Sensor: Rapid Response Zirconia Sensor / Remote Mount
Measurement Range: 0-10PPB/0-100PPB/0-1,000PPB/0-10,000PPB/ 0-100% auto range or fixed range
Display: 19 mm LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN
Response Time: <10 min. to PPB from air
Intrinsic Error (accuracy): At constant temperature: ±2% range
Over operating temperature: ±5% range
When calibrated @ target concentration @ STP: ±1% range
Operating Temperature: 5 to 45º C
Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC or 0-5VDC or 0-10VDC analog oxygen signal
Alarm Outputs: Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. One alarm/relay for sensor heater OK.
Sample Port: 1” VCR-type fittings on remote sensor module
Electrical Connections: Removable terminal blocks on back of analyzer
Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase or 12-30VDC (battery backup capability) Out
Environmental Ratings, Analyzer: Faceplate: IP66 Housing: IP20
Warranty: Electronics and sensor: One yr.
Weight: Analyzer: 1 kg