Operating Conditions: Ambient Temperature: 0º-45ºC (32º to 113ºF), Relative Humidity: 0-90% (Non-Condensing)
Range: Auto-ranging. Indicator light specifies whether the reading is in PERCENT, PPM or PPB. Display resolution to 0.05PPB.
Accuracy: Analyzing ultra-pure inert gasses with no significant reducing gas contamination (CO, H2, Hydrocarbons). The OA-1S+ accuracy specification is 2% of Range or 10PPB (whichever is greater).
Response Time:
Change response:
95% of step change in 5 seconds at 1 liter/min. sample flow rate within PPM-% levels.
90% of step change in 30 sec. at PPB levels.
Recovery time:
From room air to .1% <20 sec.
Low PPM levels <5 min.
Under 500 PPB <30 min.
Operations: Pressure= 10# to 60# PSIG. Flow rate 0.5 to 1.5LPM, 115/230VAC ±20%, 50/60Hz. (via power selector switch)
Warm-Up Time: Approximately 20 min.
Dimensions: 206mm x 229mm x 244mm
Weight: 5 kg